AL Mulla, Ismael, Lowry, John P., Serra, Pier A. and O'Neill, Robert D. (2009) Development of a voltammetric technique for monitoring brain dopamine metabolism: compensation for interference caused by DOPAC electrogenerated during homovanillic acid detection. Analyst, 134. pp. 893-898. ISSN 0003-2654
PDF
JL_voltammetric-b810227a.pdf
Download (1MB)
JL_voltammetric-b810227a.pdf
Download (1MB)
Abstract
The established stability of carbon-paste electrodes (CPEs) in brain extracellular fluid was exploited to
develop a voltammetric technique to monitor the dopamine metabolite, homovanillic acid (HVA), at 10
s intervals. At the scan rates needed for this time resolution, 3,4-dihydroxyphenylacetic acid (DOPAC),
electrogenerated as a result of HVA oxidation, was observed in the cyclic staircase voltammograms,
and this interfered with the straightforward reliable quantification of HVA. However, correction of the
HVA signal, recorded in mixtures, with currents from the DOPAC and ascorbate regions of the
voltammogram allowed the reproducible construction of well behaved HVA calibration plots. These
showed good linearity, LOD values, selectivity and stability during six days of continuous CPE
exposure to a lipid medium, which served as an in-vitro model of CPE implantation in brain tissue for
future applications.
Item Type: | Article |
---|---|
Keywords: | voltammetric technique; brain dopamine metabolism; DOPAC; homovanillic acid detection; |
Academic Unit: | Faculty of Science and Engineering > Chemistry |
Item ID: | 2062 |
Depositing User: | John Lowry |
Date Deposited: | 21 Jul 2010 14:12 |
Journal or Publication Title: | Analyst |
Publisher: | Royal Society of Chemistry |
Refereed: | Yes |
Related URLs: | |
URI: | https://mu.eprints-hosting.org/id/eprint/2062 |
Use Licence: | This item is available under a Creative Commons Attribution Non Commercial Share Alike Licence (CC BY-NC-SA). Details of this licence are available here |
Repository Staff Only (login required)
Downloads
Downloads per month over past year