Wang, Yongxiang, Markham, Charles and Deegan, Catherine (2019) Assessing the time synchronisation of EEG systems. 30th Irish Signals and Systems Conference, ISSC 2019. pp. 1-6.
Preview
cm_assessing.pdf
Download (5MB) | Preview
Abstract
This study compared the synchronisation of a
medical grade Electroencephalography (EEG) system, the
g.Tec, and a consumer grade EEG system, the Emotiv. Data was
collected from both systems using the lab streaming layer (LSL).
Both EEG systems recorded an electric signal from the surface
of a customised gel phantom. The electric signal was generated
using a solar cell which was illuminated by a monitor presenting
a sequence of black and white images. Test results show that the
g.Tec had a mean delay of 51.22 ms from the stimulus onset and
the Emotiv had a mean delay of 162.69 ms from the stimulus
onset. The result should be taken into account with future ERP
studies which will use either the EEG system and the lab
streaming layer. The design of this experiment provides a smart
way to evaluate the temporal accuracy of other EEG systems.
Item Type: | Article |
---|---|
Keywords: | EEG; phantom; synchronisation; lab streaming layer; g.Tec; Emotiv; |
Academic Unit: | Assisting Living & Learning,ALL institute Faculty of Science and Engineering > Computer Science Faculty of Science and Engineering > Research Institutes > Hamilton Institute |
Item ID: | 15550 |
Identification Number: | 10.1109/ISSC.2019.8904947 |
Depositing User: | Dr. Charles Markham |
Date Deposited: | 22 Feb 2022 14:26 |
Journal or Publication Title: | 30th Irish Signals and Systems Conference, ISSC 2019 |
Publisher: | IEEE |
Refereed: | Yes |
Related URLs: | |
URI: | https://mu.eprints-hosting.org/id/eprint/15550 |
Use Licence: | This item is available under a Creative Commons Attribution Non Commercial Share Alike Licence (CC BY-NC-SA). Details of this licence are available here |
Repository Staff Only (login required)
Downloads
Downloads per month over past year